Bronkhorst FLOW-BUS interface User Manual Page 24

  • Download
  • Add to my manuals
  • Print
  • Page
    / 26
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 23
BRONKHORST®
Page24
FLOWBUSinterface9.17.024
6 TROUBLESHOOTING
6.1 LEDINDICATIONS
Led Time Indication
Green
off Continuous Powerofforprogramnotrunning
on Continuous Normalrunning/operationmode
Short
flash
0.1secon
2.0secoff
Initializationmode
Securedparameterscanbechanged
RemoteinstalltoFLOWBUS
normal
flash
0.2secon
0.2secoff
Specialfunctionmode
Instrumentisbusyperforminganyspecialfunction.
E.g.autozeroorselftest
Red
off Continuous Noerror
Short
flash
0.1secon
2.0secoff
Nodeoccupied:
Reinstallinstrument
normal
flash
0.2secon
0.2secoff
Warningmessage.
Anerroroccurredofminorimportance.
Itwouldbewisetoinvestigatethecauseofthis.
Youarestillabletoworkwithyourinstrument.
on Continuous Criticalerrormessage.Aseriouserroroccurredintheinstrument.
Instrumentneedsservicebeforefurtherusing.
WinkMode
Green
Red
Green
Redturnbyturn
slow
wink
0.2secon
0.2secoff
Winkmode
ByacommandsendviaFLOWBUStheinstrumentcan“wink”withLed’stoindicateits
positionina(large)system
normal
wink
1.0secon
1.0secoff
Alarmindication:minimumalarm,limit/maximumalarm;powerupalarmorlimitexceededor
batchreached.
fast
wink
0.1secon
0.1secoff
Switchreleased,selectedactionstarted
Page view 23